This part of IEC 61223 applies to those components of X-RAY EQUIPMENT which
a) generate, influence the propagation of, and detect X-RADIATION, and
b) process, present and store diagnostic X-ray images in RADIOLOGICAl. INSTALLATIONS with diagnostic X-ray systems for INDIRECT RADIOSCOPY and INDIRECT RADIOGRAPHY which use XRAY IMAGE INtENSIFIERS In conunctlon with analogue anchor digital storage systems:
— closed-circuit television display systems:
— cut film cameras
— cine cameras.
This standard is a part of a senes of Particular Pubhcatlons (international standards and technical reports) which define methods of testing the constancy of operation of various subsystems of diagnostic X-v EQUIPMENT.
This standard gives methods of tests for the constancy of properties of diagnostic X-RAY EQUIPMENT as described in IEC 61223-1 (see clause 2).
This part of IEC 81223 is designed to be applicable to equipment for indirect radioscopy and indirect radiography without digital imaging devices.
1.2 Object
This standard defines
— the essential parameters which describe or affect the performance of the above components of X-RAY EQUIPMENT, and
— methods of checking that variations In measured quantities related to tnose parameters are within acceptable limits, in order to maintain adequate standards of imaging wtnlst reducing unnecessary IRRADIATION of the PATIENT
These methods are based upon assessment of radiological information using appropriate TEST
DEVICES.
The purpose of the methods s
— to establish a reference level of performance when such equipment is accepted:
— to detect or verify any significant variation in functional parameters which may then require corrective actions.
This part of IEC 61223 applies to those components of X-RAY EQUIPMENT which
a) generate, influence the propagation of, and detect X-RADIATION, and
b) process, present and store diagnostic X-ray images in RADIOLOGICAl. INSTALLATIONS with diagnostic X-ray systems for INDIRECT RADIOSCOPY and INDIRECT RADIOGRAPHY which use XRAY IMAGE INtENSIFIERS In conunctlon with analogue anchor digital storage systems:
— closed-circuit television display systems:
— cut film cameras
— cine cameras.
This standard is a part of a senes of Particular Pubhcatlons (international standards and technical reports) which define methods of testing the constancy of operation of various subsystems of diagnostic X-v EQUIPMENT.
This standard gives methods of tests for the constancy of properties of diagnostic X-RAY EQUIPMENT as described in IEC 61223-1 (see clause 2).
This part of IEC 81223 is designed to be applicable to equipment for indirect radioscopy and indirect radiography without digital imaging devices.
1.2 Object
This standard defines
— the essential parameters which describe or affect the performance of the above components of X-RAY EQUIPMENT, and
— methods of checking that variations In measured quantities related to tnose parameters are within acceptable limits, in order to maintain adequate standards of imaging wtnlst reducing unnecessary IRRADIATION of the PATIENT
These methods are based upon assessment of radiological information using appropriate TEST
DEVICES.
The purpose of the methods s
— to establish a reference level of performance when such equipment is accepted: